Medication info

Neproclean Solution (Peracetic Acid Solution 1.96%)

Prescription medication
Ingredient · strength
Peracetic Acid Solution 1.96%›Strength being verified
Manufacturer / importer
HP&NC Co., Ltd.
Appearance
A colorless and transparent liquid with a pungent unique odor, contained in a plastic bottle
Packaging
5 liters/bottle, 6 liters/bottle
Approval status
Active

MFDS item code 202502866 · Approved 2025.11.05

Summary

Main use
Prevention of contamination through cleaning and sterilization of hemodialysis equipment
Who uses it
Adults
Adult dosage
1. This medicine is to be used only with hemodialysis machines for which the dilution concentration and cleaning time have been specified to allow cleaning with peracetic acid. 2. Follow the equipment manual for the dilution concentration and cleaning time of this medicine (maximum dilution concentration: at least 0.05% peracetic acid). 3. After disinfection of the device is complete and before using the equipment, be sure to check for residues of this medicine using a hydrogen peroxide residual test strip or similar, and confirm that the hydrogen peroxide level is less than 0.5ppm.

Ingredients

Manufacturer
HP&NC Co., Ltd.

Efficacy & effects

Prevention of contamination through cleaning and sterilization of hemodialysis equipment

View approved original text

Cleansing and sterilization of hemodialysis devices are crucial in providing a safe treatment environment for patients. This process effectively removes blood residues and microorganisms that may remain in the device. Proper cleaning and sterilization reduce the risk of infection and enhance the effectiveness of dialysis treatment. Therefore, cleaning and sterilizing hemodialysis devices must always be carried out thoroughly.

Dosage and Administration

Adult dosage

1. This medicine is to be used only with hemodialysis machines for which the dilution concentration and cleaning time have been specified to allow cleaning with peracetic acid. 2. Follow the equipment manual for the dilution concentration and cleaning time of this medicine (maximum dilution concentration: at least 0.05% peracetic acid). 3. After disinfection of the device is complete and before using the equipment, be sure to check for residues of this medicine using a hydrogen peroxide residual test strip or similar, and confirm that the hydrogen peroxide level is less than 0.5ppm.

Precautions

This medication should not be administered directly to humans and is not intended for direct use by patients. Patients with allergies or sensitivities to hydrogen peroxide or peracetic acid should not use dialysis machines treated with this product. This medication must strictly follow the prescribed usage and dosage and should not be used in any other way arbitrarily. This medication is corrosive; contact with eyes or skin may cause damage or irritation, so care should be taken to avoid contact with eyes, skin, or clothing during handling. After handling, hands and skin must be thoroughly washed, and if contact with eyes or skin occurs, rinse immediately with plenty of water. All individuals using this medication should be well acquainted with the methods for preventing contamination and disinfecting medical devices, as well as the precautions for handling toxic substances. In particular, the rinsing process after dialysis is a critical step to ensure the full effectiveness of this medication, so rinsing should be performed accurately according to the equipment manual. When using this product, avoid contact with skin and eyes, and always wear appropriate protective equipment such as gloves and eye protection. The guidelines for dosage and administration must be strictly followed; violations may compromise safety and effectiveness, hence it is the responsibility of the user to manage this. If this medication is accidentally ingested, drink a large amount of water immediately to dilute it, do not induce vomiting, and consult a physician without delay. This medication can cause damage to mucous membranes; therefore, gastric lavage should not be performed. This medication should not be mixed with alkaline substances, particularly bleach, as it can lead to mucosal damage or reduced effectiveness. Store in a safe place out of reach of children and do not keep for extended periods in direct sunlight or high-temperature environments. Care should be taken to avoid contact with oxidizing agents or flammable materials. It must be stored in its original container; transferring to another container or touching the vent cap may lead to product degradation or accidents. Used medication should never be put back into the container and must be disposed of safely.

Side effects

  • Eye damage
  • skin irritation
  • mucosal damage

Not everyone experiences these, and they can vary from person to person.

Storage

This medication may be sensitive to air or moisture, so it must be stored in a tightly sealed container. It is recommended to maintain a storage temperature between 1 to 30 degrees Celsius to preserve the medication's effectiveness for a longer period.

Appearance and packaging

Appearance

A colorless and transparent liquid with a pungent unique odor, contained in a plastic bottle

Package unit

5 liters/bottle, 6 liters/bottle

Questions about this medicine

No questions yet.

For anything that needs medical judgment, consult a doctor or pharmacist.

Related searches

Search tags

Suggested searches

This drug information is based on data provided by the Korea Ministry of Food and Drug Safety.
Consult a doctor or pharmacist for accurate dosage, administration, and usage instructions.