Peracetic Acid Solution (Peracetic Acid 0.13%)
Peracetic Acid Solution 0.13%
OTC medication
HP&C Co., Ltd.
MFDS item code 202301172 · Notified 2023.04.07
Inhibition of microorganisms through chemical sterilization and disinfection of medical devices (endoscopes, etc.)
This medical device is used for chemical sterilization and disinfection, especially for the disinfection of heat-sensitive endoscopes.
1. Preparation Prepare the disinfectant solution for this medicine as follows. Mix 4 L of Solution 1 with 90 mL of Solution 2. 2. Instructions for use A. Thoroughly pre-clean the medical instruments, then completely immerse them in the prepared disinfectant solution. For instruments and equipment with fine lumens or complex structures, use pressurized injection or aspiration to ensure they are thoroughly wetted with the prepared solution. Perform a leak test if necessary before use. B. Immerse completely for at least 5 minutes. To achieve spore-killing, immerse completely for at least 10 minutes. C. After thorough immersion, remove the instruments and rinse them under running sterile purified water for at least 15 seconds. A larger volume of water may be used depending on the intended use. Since disinfectant solution can easily remain inside instruments with fine lumens or complex structures, thoroughly rinse them by injecting sterile purified water or water under pressure, or by increasing the rinsing time. 3. After disinfecting the medical instruments and before using them, be sure to check for residues of this medicine using a hydrogen peroxide residue test strip or similar, and confirm that the hydrogen peroxide concentration is less than 0.5 ppm.
This product must not be used directly on the human body. If it comes into contact with skin, immediately remove contaminated clothing and wash thoroughly with running water. If pain persists, consult a dermatologist. If it gets into the eyes, immediately rinse with running water for more than 15 minutes and seek treatment from an ophthalmologist. Delays or insufficient rinsing may cause eye damage. If inhaled, move to an area with fresh air and seek consultation with a specialist. If swallowed accidentally, drink a large amount of water or milk, do not induce vomiting, and seek medical attention. Inducing vomiting can cause respiratory problems. When using this drug, use a container with a lid to prevent steam from spreading, and use it in a well-ventilated area. If necessary, it is advisable to use it in a place with ventilation facilities. Wear protective equipment such as goggles to avoid eye contact, and if it does get into the eyes, immediately rinse with a large amount of water and seek treatment from a specialist. This medication is effective only when the concentration of peracetic acid is above 0.05%, so check the concentration before use. If there are contaminants such as blood or body fluids on the instruments to be disinfected, the effectiveness of the solution may be reduced, so instruments must be cleaned thoroughly before disinfection. Especially for complex instruments like endoscopes, follow the manufacturer's cleaning instructions. After cleaning, thoroughly remove moisture from the instruments before immersing them in the disinfectant, and after disinfection, check for any residual peracetic acid; if residual acid is present, extend the rinsing time to completely remove it. To kill spores, the instruments must be submerged for more than 10 minutes, and some instruments may rust or degrade, so do not use continuously for more than 1 hour. Medical devices made of natural rubber or latex may deteriorate upon repeated use, so check before use. If the instruments are cracked or rusted, do not use them as the disinfection effect will be reduced. During disinfection, wear protective equipment such as rubber gloves, gowns, masks, and goggles to prevent contact with infectious substances or solutions on the skin or by inhalation. Store this medication out of reach of children and avoid accidental ingestion or swallowing. The prepared disinfectant solution should be stored in a sealed container away from direct sunlight and at room temperature, and the lid should be tightly closed during use. Ensure that the concentration of peracetic acid in the disinfectant solution is above 0.05% before use, and be aware that moisture or organic matter may reduce its effectiveness. This product is acidic, so mixing it with chlorine compounds such as hypochlorite can generate hazardous chlorine gas; do not mix. When disposing, dilute with a large amount of water, and wear gloves, goggles, and masks during disposal to avoid contact with the solution. Disposal methods include neutralizing by mixing with an alkali or adding a reducing agent and then disposing of it. Store this product tightly closed in its original container; transferring it to another container may pose a risk of accidents or quality degradation.
Not everyone experiences these, and they can vary from person to person.
This product should be stored in a sealed container to protect it from air and moisture. It is best to maintain a storage temperature between 1 and 30 degrees Celsius to preserve its effectiveness.
Component 1: A colorless, transparent liquid with a peculiar odor Component 2: A colorless, transparent liquid with a slight orange scent
Component 1 - 4 liters/bottle, Component 2 - 90 milliliters/bottle
Even with the same ingredient, strength, dosage form and how it is taken can differ.
Even with the same ingredient, the dose and directions can differ by strength and dosage form. Talk to a doctor or pharmacist before switching from a medication you are already taking.