Medication info

Endopac (Glacial Acetic Acid 0.2%)

OTC medication
Ingredient · strength
Peracetic Acid Solution 0.2%›Strength being verified
Manufacturer / importer
Person Co., Ltd.
Appearance
(Component 1) Colorless transparent liquid with the smell of acetic acid (Component 2) Colorless to pale yellow transparent liquid with the smell of alcohol
Packaging
(Component 1) 4 L/plastic bottle, (Component 2) 90 mL/plastic bottle (Component 1) 5 L/plastic bottle, (Component 2) 112.5 mL/plastic bottle
Approval status
Active

MFDS item code 201607103 · Approved 2016.12.14

Summary

Main use
Suppression of microorganisms through sterilization and disinfection of medical devices
Who uses it
Adults
Adult dosage
1. Preparation Prepare the disinfectant solution according to the following method. Mix Solution 1 and Solution 2 in a ratio of 400:9. 2. Directions for use A. Completely immerse the pre-cleaned medical instruments in the prepared disinfectant solution. For instruments and equipment with narrow lumens or complex structures, use pressure injection or suction to ensure thorough contact with the prepared disinfectant solution. If necessary, perform a leak test before use. B. Keep completely immersed for at least 5 minutes. For spore inactivation, keep completely immersed for at least 10 minutes. C. After thorough immersion, remove the instruments and rinse them under running sterile purified water for at least 15 seconds. Depending on the intended use, a larger volume of water may be used. Since disinfectant solution can readily remain inside instruments with narrow lumens or complex structures, rinse thoroughly by pressure-injecting sterile purified water or water, or by extending the rinsing time. 3. After disinfection of the medical instruments is complete and before using them, be sure to check for residual medicine using a hydrogen peroxide residual test strip or similar device and confirm that the residual hydrogen peroxide concentration is less than 0.5 ppm.

Ingredients

Manufacturer
Person Co., Ltd.

Efficacy & effects

Suppression of microorganisms through sterilization and disinfection of medical devices

View approved original text

This product is used for the sterilization and disinfection of medical devices. In the sterilization process, all microorganisms present on the medical devices are eliminated to prevent infection, while in the disinfection process, harmful bacteria are effectively removed to maintain a safe medical environment. This helps reduce the risk of infection for patients and supports safe treatment.

Dosage and Administration

Adult dosage

1. Preparation Prepare the disinfectant solution according to the following method. Mix Solution 1 and Solution 2 in a ratio of 400:9. 2. Directions for use A. Completely immerse the pre-cleaned medical instruments in the prepared disinfectant solution. For instruments and equipment with narrow lumens or complex structures, use pressure injection or suction to ensure thorough contact with the prepared disinfectant solution. If necessary, perform a leak test before use. B. Keep completely immersed for at least 5 minutes. For spore inactivation, keep completely immersed for at least 10 minutes. C. After thorough immersion, remove the instruments and rinse them under running sterile purified water for at least 15 seconds. Depending on the intended use, a larger volume of water may be used. Since disinfectant solution can readily remain inside instruments with narrow lumens or complex structures, rinse thoroughly by pressure-injecting sterile purified water or water, or by extending the rinsing time. 3. After disinfection of the medical instruments is complete and before using them, be sure to check for residual medicine using a hydrogen peroxide residual test strip or similar device and confirm that the residual hydrogen peroxide concentration is less than 0.5 ppm.

Precautions

This drug must not be used directly on the human body. If it comes in contact with skin, contaminated clothing should be removed immediately, and the affected skin should be washed thoroughly with running water. If pain persists, consult a dermatologist. If it enters the eyes, rinse immediately with running water for at least 15 minutes and consult an ophthalmologist; delaying or insufficient rinsing may cause eye damage. If vapors are inhaled, move to an area with fresh air and consult a specialist. If swallowed accidentally, drink plenty of water or milk, do not induce vomiting, and seek medical attention. When using this product, store it in a container with a lid to prevent vapor dispersion and ensure proper ventilation; use in an area with ventilation if necessary. Always wear protective eyewear to avoid contact with eyes. Ensure that the concentration of glacial acetic acid is above 0.05%, and if there are contaminants such as blood or bodily fluids on the device, the disinfection effect may be reduced, so wash thoroughly before use. After washing, remove all moisture before immersing in the disinfectant. Check for residual glacial acetic acid in the disinfectant; if present, extend the rinsing time to ensure all residues are removed. To kill spores, immerse for more than 10 minutes, but do not immerse some instruments for more than 1 hour. Instruments made of natural rubber or latex may be damaged with repeated use, so check carefully, and do not use on cracked or rusted instruments. When disinfecting, wear protective equipment such as rubber gloves, gowns, masks, and protective eyewear to prevent contact with the disinfectant and minimize infection risk. During storage, keep it out of reach of children, place it back in its original container, close the lid tightly, avoid direct sunlight, and store at room temperature. Close the lid during use to prevent vapor dispersion. This product is acidic, so do not mix it with chlorine-based products as it can produce dangerous chlorine gas. When disposing, dilute with plenty of water, and it is advisable to neutralize with alkali or reducing agents to prevent environmental pollution. Always wear safety equipment when handling this product to avoid direct contact with skin or eyes.

Side effects

  • Pain
  • eye disorders
  • respiratory system disorders

Not everyone experiences these, and they can vary from person to person.

Interactions

Drug interactions

Do not use this drug in combination with chlorine-based agents such as hypochlorite.

Storage

This drug may be sensitive to air or moisture, so it must be stored in a tightly sealed container. The storage temperature should be maintained between 1 degree and 30 degrees Celsius at room temperature for safe preservation of the drug's effectiveness.

Appearance and packaging

Appearance

(Component 1) Colorless transparent liquid with the smell of acetic acid (Component 2) Colorless to pale yellow transparent liquid with the smell of alcohol

Package unit

(Component 1) 4 L/plastic bottle, (Component 2) 90 mL/plastic bottle (Component 1) 5 L/plastic bottle, (Component 2) 112.5 mL/plastic bottle

Same-ingredient medicines

Even with the same ingredient, strength, dosage form and how it is taken can differ.

Even with the same ingredient, the dose and directions can differ by strength and dosage form. Talk to a doctor or pharmacist before switching from a medication you are already taking.

Questions about this medicine

No questions yet.

For anything that needs medical judgment, consult a doctor or pharmacist.

Related searches

Search tags

Suggested searches

This drug information is based on data provided by the Korea Ministry of Food and Drug Safety.
Consult a doctor or pharmacist for accurate dosage, administration, and usage instructions.